AAMI
Electrode Test Results
Wet
Gel Electrode Tests
6%
Potassium Chloride Gel
The
following products incorporate 6% Potassium Chloride Gel
and are tested according to the voluntary Association for
the Advancement of Medical Instrumentation (AAMI) standard
requirements for electrical performance for disposable ECG
electrodes:
|
LL
Series
|
LRM
Series
|
FO
Series
|
LLEBL
Series
|
LLEBX
Series
|
P-6
|
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.15
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
51
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
0.7
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
49
|
Not
more than 3000
|
10%
Potassium Chloride Gel
The
following product incorporates 10% Potassium Chloride Gel
and is tested according to the voluntary Association for
the Advancement of Medical Instrumentation (AAMI) standard
requirements for electrical performance for disposable ECG
electrodes:
- Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.15
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
50
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
1.1
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
47
|
Not
more than 3000
|
Solid Gel Electrodes
All
Lead-Lok solid gel electrodes are tested according to
the voluntary Association for the Advancement of Medical
Instrumentation (AAMI) standard requirements for electrical
performance for disposable ECG electrodes and exceed those
performance standards.
Stealth
Gel Electrodes
The
following products are injected with Stealth Gel, Lead-Lok's
proprietary solid gel:
|
SFT
Series
|
SFS
Series
|
SFL
Series
|
SFLL
Series
|
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.12
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
60
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
6.5
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
46
|
Not
more than 3000
|
Pre-Wired
Electrodes
The
following pre-wired electrodes are laminated with a unique
conductive hydrogel:
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.51
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
237
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
12.8
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
224
|
Not
more than 3000
|
The
following snap electrodes are laminated with a unique
conductive hydrogel:
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.13
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
207
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
12.5
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
220
|
Not
more than 3000
|
Pediatric Electrotherapy Electrodes
All
Lead-Lok electrotherapy electrodes are tested according
to the voluntary Association for the Advancement of Medical
Instrumentation (AAMI) standard requirements for electrical
performance for disposable ECG electrodes and exceed those
performance standards.
Pre-Wired
Electrodes
The
following pre-wired electrodes are laminated with a unique
conductive hydrogel:
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.51
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
237
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
12.8
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
224
|
Not
more than 3000
|
Snap
Hydrogel Electrodes
The
following snap electrodes are laminated with a unique conductive
hydrogel:
Test
Results:
| |
Lead-Lok
Typical
|
AAMI
Limits
|
| DC
Offset Voltage (mV) |
0.13
|
Not
more than 100
|
| AC
Impedance @ 10Hz (OHMS) |
207
|
Not
more than 2000
|
| DC
Offset @ 5 sec after capacitor discharge (mV) |
12.5
|
Not
more than 100
|
| Recovery
Slope @ 30 second interval (mV/sec) |
0.0
|
Not
more than 1.0
|
| AC
Impedance after defib (OHMS) |
220
|
Not
more than 3000
|